Thermo Scientific Miran Sapphire 205B is a family of portable, lightweight infrared instruments that can monitor ambient air in workplace environments for the purpose of detecting the presence of gases. Thermo Scientific Miran Sapphire 205B exclusive design of the wavelength generator filter allows accurate and fast wavelength selection separating the MIRAN SapphIRe Analyzer from other portable gas analyzers.
Thermo Scientific Miran Sapphire 205B contains a single-beam infrared spectrophotometer to specifically and accurately measure many gases with a single unit. Thermo Scientific Miran Sapphire 205B results of this analysis can be displayed in parts per billion (ppb), parts per million (ppm), percent (%), mg/m3, or absorbance units
Thermo Scientific Miran Sapphire 205B models can be configured with two options: intrinsic safety certification and spectral scanning using Thermo Scientific ThermoMatch Spectrum Correlation Software.
Thermo Scientific Miran Sapphire 205B Applications:
- Industrial hygiene monitoring
- Emergency responce analysis
- Indoor air quality studies
- Hospital gas monitoring
- Fume hood/Tracer gas testing
- Leak detection
Thermo Scientific Miran Sapphire 205B Features:
- Exclusive wavelength generator filter design
- Library of up to 121 components
- Sensitive for sub-ppm detection
- Versatile and upgradable
- Lightweight and portable
Thermo Scientific Miran Sapphire 205B Included:
- 2 Battery Packs (8 hours total run)
- Charger Cables
- RS-232 cable
- Thermoconnect software
- Manual
You may also be interested in Thermo Scientific TVA 1000B Toxic Vapor Analyzer
Thermo Scientific Miran Sapphire 205B Specifications:
Measurement Range | See Appendix H for each compound range. |
Concentration | Compound Dependent |
Absorbance | 0.000 to 1.2 Au (practical upper limit) |
Measurable Gases | SapphIRe-DL: up to 5 gases in a background of 5 additional gases. Storage for up to 10 single gas and 10 multigas applications. SapphIRe-SL: same as SapphIRe-DL, except factory calibrated for 50 single gases. SapphIRe-XL: same as SapphIRe-SL, except factory calibrated for 120 single gases. |
Output | Digital readout of concentration (ppb, ppm, %, mg/m3 ) or absorbance units (AU). Also connections for analog output, relay contacts, parallel port (printer) and serial communication (RS232). |
Wavelength | Seven fixed bandpass filters and one continuously linear variable filter (7.7 to 14.1 microns). The seven fixed wavelengths are: 1.859 water vapor, 3.333 general hydrocarbons, 3.573 formaldehyde, 4.004 reference, 4.250 carbon dioxide, 4.500 nitrous oxide, and 4.620 carbon monoxide. |
Pathlength | Two user-selected settings: 0.5 and 12.5 meters nominal |
Battery Pack | Type: Internal, rechargeable, sealed, intrinsically safe NiCd (NickelCadmium) battery pack. Voltage: 7.2 V nominal. Capacity: 5.7 Ah. |
Battery Operating Time | Up to four hours of continuous operation at 23°C. |
Battery Recharge Time | Four hours after complete discharge if analyzer is off. Approximately two hour for each hour of discharge from full capacity if analyzer of off. |
Operator Error Alarm | User definable: none, double beep, error screen. |
Concentration Alarm | Low, High, and STEL |
Power Requirements of ac/dc Converter | 95 to 260 V ac at 50/60 Hz, ±3% or 11 to 15 V dc, 5 A (from automobile battery with optional dc power cable) |
Measurement Cycle Time | Single gas analysis screen updated each half-second. Dependent upon application in multigas mode or if compensations are enabled. |
Reference, Operating, and Storage Conditions | See Table 2-2. |
Electromagnetic Compatibility (EMC) | Complies with the requirements of the European EMC Directive 89/336/EEC. |
Sample Flow Rate | 14 liters/minute (nominal) with sampling wand and particulate filter. |
Sample Cell Volume | 2.23 liters |
Gas Purging | 5 volumes of sample required to purge cell by 99%. |